Diamond-based radiation detectors |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Model: |
ADII-1.1 |
ADII-1.2 |
ADII-1.3 |
| Energy resolution* |
≤2% |
2-5% |
5-10 % |
| Deviation of energy resolution value, less than* |
2% |
2% |
2 % |
| Displacement of energy peak, less than* |
5 % |
5 % |
5 % |
| Non-stability of counting, less than* |
2 % |
2 % |
2 % |
| Thickness of semiconductor element |
0,1..0,4 mm |
0,1..0,4 mm |
0,1..0,4 mm |
| Surface area of semiconductor element |
2..9 mm2 |
2..9 mm2 |
2..9 mm2 |
*These are standard option QC and measured under working voltage for alpha-particles with energy of 5,5MeV for 8 hours period
Other characteristics
| Model: |
ADII-1.1 |
ADII-1.2 |
ADII-1.3 |
| Bias voltage |
100..400 V |
100..400 V |
100..400 V |
| Dark current, less than |
0,5×10-12 A |
0,5×10-12 A |
0,5×10-12 A |
| Impulse rise time, less than |
1,0×10-9 sec |
1,0×10-9 sec |
1,0×10-9 sec |
Optional casing:
Mounting into a case as provided by customer
Mounting into a case designed according to customer's specifications
Optional testing
| Radiation type: |
Neutrons |
Protons |
Heavy Ions |
| Energy resolution |
|
|
|
| Deviation of energy resolution value |
|
|
|
| Displacement of energy peak |
|
|
|
| Non-stability of counting for 8 hours |
|
|
|
QC Test Results Datasheet hard copy (in electronic format upon request)
ADII-2 detector has been designed to make measurements of volume and specific activities of radionuclides in liquid or gas environment. The detector can work either in counting or current mode. The detector has been successfully used for dose monitoring of photon and high-energy corpuscular radiation.
Standard Option: 15mm x 27mm metal case
Quality Control Certificate
+ QC Test Results Datasheet hard copy (in electronic format upon request)
Standard (Beta-radiation) QC Testing as in the Table below
| Sensitivity, not less |
1,4×10-19 C×mm2×b-part.-1 |
| Dark current, less than |
0,5×10-12 A |
| Non-stability of counting for 8 hour, less than |
2 % |
| Thickness of semiconductor element |
0,1..0,6 mm |
| Surface area of semiconductor element |
2..10 mm2 |
Other characteristics
| Energy range: Alpha-particle Beta-particle Gamma & X-Ray radiation |
4,5 MeV to 7,6 MeV; 65 keV to 1,0 MeV. 2 keV to 3 MeV |
| Measuring range |
104 ..1010 becquerel/dm3 |
| Impulse rise time, less than |
1,0×10-9 sec |
| Bias voltage |
50..200 V |
| Sensitive element |
Natural diamond |
Optional casing:
Chemically resistant 12mm x 50mm polystyrene case with cable
Mounting into a case as provided by customer
Mounting into a case designed according to customer's specifications
Optional testing
|
Radiation type: |
Alpha-particle |
Gamma-radiation |
X-ray radiation |
|
Sensitivity |
|
|
|
|
Non-stability of counting for 8 hours |
|
|
|
QC Test Results Datasheet hard copy (in electronic format upon request)
ADII-3 line of detectors has been designed to make dose measurements of photon, electron and proton radiations and to be used in dosimetric devices. Its high sensitivity and small active volume allow the detector to be successfully used to measure radiation characteristics of therapeutic radiation beams, radiation fields and phantom absorbed doses. Also, the detector can be successfully used for direct in-vivo measurements of radiation fields and absorbed doses on human patients. The detector is compatible with Max 4000, PTW UNIDOS, CNMC1150/1100, 292 Therarad, Keithley 6517 electrometers.
Standard Option: waterproof 7mm x 20mm polystyrene case with cable
Quality Control Certificate
+ QC Test Results Datasheet hard copy (in electronic format upon request)
Standard QC Testing as in the Table below
| Model: |
ADII-3.1 |
ADII-3.2 |
ADII-3.3 |
| Linearity, 0.05 Gy/min to 5.0Gy/min |
≤2% |
2..4% |
4..6 % |
| Sensitivity |
20..300 nC/Gy |
20..300 nC/Gy |
20..300 nC/Gy |
| Dark current, less than |
0,5×10-12 A |
0,5×10-12 A |
0,5×10-12 A |
| Thickness of semiconductor element |
0,1..0,4 mm |
0,1..0,4 mm |
0,1..0,4 mm |
| Sensitive volume |
1..6 mm3 |
1..6 mm3 |
1..6 mm3 |
Other characteristics
| Model: |
ADII-3.1 |
ADII-3.2 |
ADII-3.3 |
| Range of dose rate |
0,05..30 Gy/min |
0,05..30 Gy/min |
0,05..30 Gy/min |
| Energy range: Photons Electrons |
0,08..25
MeV |
0,08..25
MeV |
0,08..25
MeV |
| Pre-irradiation dose, less than |
10 Gy |
10 Gy |
10 Gy |
| Maximum adsorption dose, not less |
107 Gy |
107 Gy |
107 Gy |
| Bias voltage |
50..200 V |
50..200 V |
50..200 V |
| Sensitive element |
Natural diamond |
Natural diamond |
Natural diamond |
Optional casing:
15mm x 27mm metal case
Mounting into a case as provided by customer
Mounting into a case designed according to customer's specifications
Recommended operation conditions: the temperature of an environment
from +5°C to +30°C; relative humidity up to 85 % at normal temperature.
Available
types of cases are: Hermetic water-proof polystyrene case;
Metal cases based on standard BNC connector.